Operation Method
Collection of Measured Amount of Distilled Water
In case that the tank becomes empty while collecting water
FILL
CONDUCTIVITY
M
F
WATER
LEVEL
MEASURED
CONTINUOUS
FULL
EMPTY
POWER
PURE
WATER
DISTILLED
WATER
In Distillation
Tank Empty
Distill-Water NA
Measured: 15.0L
1. The left screen is displayed when the tank
becomes empty during collecting water, and
DISTILLED WATER lamp blinks and operation of
water collection is suspended.
2. Distillation starts.
3. In case that "EMPTY" lamp of WATER LEVEL
display
does
not
light
up,
press
DISTILLED WATER key again to start collecting
water. If PURE WATER key is pressed under this
condition, the holding condition is released.
4. After collecting water is finished, "Measured: 0.0L"
appears on the window, and return to the screen of
Clause 2.
Cautions
1. During distillation and ion exchange, the supply water circulation solenoid pump is always running.
When the distilled water tank is full, this pump stops. To sample pure water while this pump is
stopped, set this pump normally ON in the sub-menu and sample pure water after stabilization of
the water quality. This pump can be set to normal operation from the sub-menu while the distillation
water tank is full.
2. If the supply water circulation pump has been set to normally ON, its setting is cancelled under
following conditions:
①
Power key turned OFF
②
Distilled water sampled, the storage tank water level lowering, and distillation resumed
③
Abnormality