2. Before Using this unit
Requirements for Installation
WARNING!
20. Install firmly the ion exchange resin cartridge (CPC-N) and high-purity cartridge
(CPC-H).
Install the ion exchange resin cartridge (CPC-N) and high-purity cartridge (CPC-H) following
the procedure shown below.
Connect securely because insecure connection may cause water leakage.
1. Make sure that the earth leakage breaker of this unit is "Off" and that the tap is tightened.
2. Take the ion exchange resin cartridge and high-purity cartridge out of attachments to the
unit.
3. Place the ion exchange resin cartridge and high-purity cartridge taken out on the receiver
within the unit. (See Fig.1.)
4. Fix the ion exchange resin cartridge and high-purity cartridge with the band of receiver.
(See Fig.2.)
Fig.1
Fig.2
5. Remove the rubber cap attached to the inlets and outlets of ion exchange resin cartridge
and high-purity cartridge.
6. Fit in the coupler marked with (IN) to the inlet of ion exchange resin cartridge (left) until click
is heard and repeat the procedure in the same way for high-purity cartridge. (See Fig.3.)
7. Fit in the coupler marked with (OUT) to the outlet of ion exchange resin cartridge (right)
until click is heard and repeat the procedure in the same way for high-purity cartridge. (See
Fig.3.)
IN
リセットスイッチ
IN
OUT
OUT
Fig.3
Coupler may be hard at first. When applying force in inserting, do not make it curved
because insertion port may be broken.
8. Coupler can be removed easily from the ion exchange resin cartridge by pulling it toward
yourself while pushing the black part of the coupler to the depth.