XO FLOW – Installation instructions
9
Suction pipe
Plastic pipe Ø 40 mm with socket – see YC-001.
6 cm max.
Incoming
(compressed) air
Pipe 3/8” female internal thread – preferably fitted with a ball
valve – see YC-001.
Incoming air:
•
Air pressure 5.5 – 7,5 bar
•
Air flow rate > 55 l/min
•
Humidity dew point < -20°C at atmospheric pressure
•
Oil contamination max. 0.5 mg/m
3
•
Particulate contamination < 100 particles/m
3
(particle size 1 –
5 µm)
If the incoming air pressure exceeds 7
,
5 bar, a reduction valve must be
fitted.
Air quality must be in accordance with local air quality
regulations.
7 cm max.
Incoming water
Pipe 3/8” female internal thread – preferably fitted with a ball
valve – see YC-001.
Incoming water:
•
Inlet pressure 2.5 – 6 bar
•
Water flow rate > 5 l/min
•
pH: 6.5 – 8.5
•
Maximum particle size < 100 µm
If the incoming water pressure exceeds 6 bar, a reduction valve must be
mounted in front of the unit.
Water quality must be in accordance with local drinking water
regulations.
Maximum inlet water conductivity: 850 µS/m
7 cm max.
Wastewater
Plastic pipe Ø 32 mm with socket – see YC-001.
Gradient of wastewater lines > 1%
Drainage capacity > 10 l/min
6 cm max.
To avoid the risk of electric shock, this equipment must be connected to a supply mains
with protective earth.
To avoid the risk of electric shock always switch off the mains switch before opening or
touching the internal components.
If the incoming water pressure exceeds 6 bar, a reduction valve must be mounted in front
of the unit.
To ensure correct operation of the unit, the compressor must be an oil-free type and fitted
with an air dryer.
If the incoming air pressure exceeds 8 bar, an appropriate reduction valve must be fitted.
2.4
Operating conditions
XO FLOW must be operated under the conditions shown below:
Table 2 – Operating conditions
Condition
Operation
Temperature
+15
o
C – +35
o
C
Relative humidity
20% – 95%
Air pressure
700 hPa – 1060 hPa