XO FLOW – Installation instructions
8
Using the supplied expansion anchors and screws, the unit shall preferably be fixed to a concrete floor
with the following minimum characteristics:
•
Concrete class C20/25 to C50/60, not cracked
•
Mounting bolts to be fixed at least 120 mm from the edge or other openings in the concrete
If the operatory has a wooden floor you must make sure that the floor construction is stable enough to
withstand the forces applied from the unit to the floor.
If it is not possible to drill holes in the floor or if the floor is not stable enough, the unit can
be installed on an installation plate – see section 5.2.7.
2.3
Supplies and connections
Before installing the unit, please make sure that the installation supply requirements listed in Table 1
are available.
Table 1 – Installation requirements
Room
Requirement
Width
Minimum in combination with XO WORKTOP = 200 cm (recommended 220 cm)
Length
Minimum in combination with XO WORKTOP = 320 cm (recommended 340 cm).
Ceiling height
> 230 cm (recommended 255 cm)
Electrical & IT
Requirement
Length
above floor
surface
Mains supply
230 VAC + 10%, 50 Hz. Min. 3 x 1.5 mm
2
PVC cable with earth
rated for >75°
40 cm
Main fuse
The electrical installation must be secured with a 10 A fuse.
Equipotential
earth (if required
by national law)
1 x 4.0 mm
2
40 cm
Assistant call
control cable
Max. 5A /30V DC or 3A/250V
50 cm
X-ray unit
attached to the
XO unit
Cable for X-ray must have its own installations pipe. Cable shall
be connected to an installation box in the floor. Cables must be
shielded and conform to the requirements in EN IEC 60601-1 and
EN IEC 60601-1-2.
150 cm
Suction motor
control cable
Max 5A /30V DC or 3A/250V AC.
50 cm
Ethernet cable
The unit may be connected to a network via a CAT6 S/FTP
shielded ethernet cable.
70 cm
USB cable
“Instructions for use, XO FLOW”
50 cm
Positioning of
cables in the floor
See installation drawing YC-001.
Suction, air and
water
Requirement
Height
above floor
surface
Suction
Suction machine power >600 l/min.
Vacuum pressure at the connection point under static conditions:
Min = -35 mbar, Max = -150 mbar.