XO FLOW – Installation instructions
67
Be sure that the balance suspension for the scaler is adjusted so that the hose remains in the
instrument-disinfection-holder while the unit water lines are disinfected.
6.
Test that all “buttons” on the foot control function correctly.
7.
Test the operating light.
8.
Synchronize the patient chair using the technical settings app.
9.
Check that the patient chair moves up/down and that the back rest can be adjusted correctly.
10.
11.
Test the patient chair collision stop.
12.
If installed, test the assistant call function.
13.
Check the network connection – and if required also that the unit is connected to the internet.
14.
Tap
to shut the unit down and test the suction disinfection system and the water
disinfection system.
15.
Check the unit for leakages in water, air, suction and drain systems.
Fit the unit stand panels carefully.
Remember to connect the protective earth wires to each panel.
After the panels are fitted:
1.
Move the patient chair up and down and check that the clearance between the panels and the
patient chair support is uniform in the whole range of movement.
2.
Check that all 3 cover plugs are installed correctly in the unit’s column.
3.
Check the unit for scratches or other handling damages.
8
INSTALLATION REPORT
Fill in the installation report and submit it to
Access the installation report here (Partner login area)