XO FLOW – Installation instructions
55
Figure 77 – Patient chair support
Tighten the nuts thoroughly with a 19 mm top wrench. See Figure 78.
Figure 78 – Fitting the patient chair
Route the chair cables through the hole in the patient chair support and attach the power
cable (1) to the PCB connector X1, communication cable (2) to the PCB connector X3A, the
suction holder cable (3) to the suction holder and the ground wire (4) to the motor. See
Figure