47
XCell
TM
ATF System with C410:V3 Controller
User Guide
•
Vacuum source connected
6.3.1 Following LOGON and entry of BATCH Information, go to the SET UP screen.
6.3.2 Go to the BASIC SET UP screen, and enter the following field values.
Field
VALUE
Controller Set UP
Pump Model No
Select ATF 6
Control Mode
Select FLOW
Slope Function Enable
Don’t enable
Bioreactor pressure
0
∆
height (cm)
0
Alarm Delay(min)
1
Alarm Set Points
VALUE
HiHi Alarm Set Points
7 psi
LoLo Alarm Set Point
-7 psi
Alarm Delay (x10 msec)
50 for both
Pump Parameters
VALUE
P-Pressure Offset (psi)
0.5
E-Pressure Offset (psi)
-0.5
P-Delay (%)
70
E-Delay (%)
70
P-Overtime (%)
120
E-Overtime (%)
120
P-PV Step Size (%)
3
E-PV Step Size (%)
3
7.3.3 Press Accept Changes.
7.3.4 Go to the ADVANCED SET UP screen and enter the following field values
Field
VALUES
User Set Point Range
XCell
TM
ATF4
XCell
TM
ATF6
XCell
TM
ATF8
XCell
TM
ATF10 lgcy
XCell
TM
ATF10
Max P-FLOW (LPM)
10
20
40
80
100
Min P-FLOW (LPM)
1
5
10
20
20
Max E-FLOW (LPM)
10
20
40
80
100
Min E-FLOW (LPM)
1
5
10
20
20
PV1 Max (%)
95
PV1 Min (%)
5
PV2 Max (%)
95
PV2 Min (%)
5
Output (CV) Limits
VALUE
PRV1 Max (psi)
25
PRV1 Min (psi)
1