19
XCell
TM
ATF System with C410:V3 Controller
User Guide
k.
Interconnect Cable Plug
l.
Operator Interface Terminal (OIT or HMI)
m.
Illuminated POWER ON/OFF Selected Switch
n.
Illuminated System STOP Push Button
o.
Alarm Relay Outputs (2)
p.
Harvest Pump Relay Outputs (2)
q.
Ethernet Plug
Figure 6.
Electric Box (E-Box) Connections
5.2
Primary Pneumatic Services
Air inlet (e) -
Located on the P-Box and provides an inlet to house compressed air
source. Recommended minimum air pressure requirement is ~50psi / 3.4bar.
Somewhat higher inlet pressures may be required, as needed, to generate higher flow
rates or to drive pneumatic instruments.
Do not exceed 90 psi / 6.1 bar on the Air inlet
Exhaust/ vacuum line (f) -
Located on the P-Box.
This line is connected to a vacuum
source. A vacuum source should always be connected to the exhaust line to ensure
reliable operation of the XCell
TM
ATF.