OPERATING MANUAL

ba75507e05 07/2016

Turb 430 IR/T

HANDHELD TURBIDIMETER

Summary of Contents for Turb 430 IR/T

Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...

Page 2: ...ba75507e05 07 2016 Turb 430 IR T Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...etting the language 24 4 3 4 Navigation example 2 setting the date and time 25 4 3 5 Menu overview 26 4 4 System settings System menu 28 4 4 1 Measured value memory 29 4 4 2 Display 31 4 4 3 Interface...

Page 4: ...1 5 1 1 Inserting exchanging the batteries 51 5 1 2 Retrofitting the accumulator pack 52 5 2 Cleaning 53 5 2 1 Cleaning the cell shaft 53 5 2 2 Cleaning the cells 53 5 3 Disposal 54 6 What to do if 55...

Page 5: ...dheld meter provides the maximum degree of operating comfort reliability and measuring certainty for all applications Note If you need further information or application notes you can obtain the follo...

Page 6: ...entries start measurement START ENTER Call up the Configuration menu all settings are made here MENU Switch the measuring instrument on off EIN AUS Output display contents to RS232 interface e g print...

Page 7: ...ia the number keys 1 3 Display The graphic display shows all information of the current measurement in the measured value display The illumination enables to read the display even in the darkness Exam...

Page 8: ...tly use the Turb 430 IR T in the laboratory Laboratory operation with the LabStation enables the following additional functions Line power operation is possible to save the batteries or accumulator pa...

Page 9: ...ns to take when handling chemicals The personnel responsible for the commissioning operation and maintenance must have the necessary qualifications for this work If the personnel do not have the requi...

Page 10: ...SPOSAL Non compliance results in the loss of warranty claims Follow the points listed below when operating the photometer Follow the local safety and accident prevention regulations Observe the enclos...

Page 11: ...gthy period of time is visibly damaged no longer operates as described in this manual If you are in any doubt please contact the supplier of the instrument CAUTION Danger of eye damage by visible and...

Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...

Page 13: ...optional parts of the scope of delivery are available as accessories see section 8 1 3 2 Power supply You can operate the meter either with batteries accumulator pack or a power pack The power pack s...

Page 14: ...tteries or accumulator pack the meter automatically switches off the display illumination if no key is pressed for 30 seconds The illumination is switched on with the next keystroke again The display...

Page 15: ...tion and place the Turb 430 IR T in the LabStation Connecting the LabStation optional 1 If necessary replace the Euro plug 1 on the power pack 2 by the country specific plug suitable for your country...

Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...

Page 17: ...pack see section 3 2 operation with LabStation connect the LabStation and place the Turb 430 IR T in the LabStation see section 3 3 Switch on the meter see section 4 1 Set the language as necessary se...

Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...

Page 19: ...automatically switches to the measuring mode Switching off Press the ON OFF key Automatic switchoff function The meter has an automatic switch off function in order to save the batteries or accumulat...

Page 20: ...or 30 seconds The illumination is switched on again with the next keystroke 4 2 Inserting a cell To be able to insert cells in the Turb 430 IR T the cell shaft has to be prepared to take in a cell Ins...

Page 21: ...e display indicates measurement data in the measured value display Calibration The display indicates a calibration procedure with calibration information Data transmission The meter transmits measurin...

Page 22: ...esignated by the name of the function They are immediately carried out by confirming with START ENTER Example display the Calibration record function in the Turbidity menu Messages Information or oper...

Page 23: ...ained in the two following sections by reference of examples Setting the language section 4 3 3 Setting the date and time section 4 3 4 i Insert standard 1000 FNU NTU i Press and hold START i Align sa...

Page 24: ...tion menu with MENU The instrument is in the configuration mode 2 Select the System menu with The current selection is displayed in reverse video 3 Open the System menu with START ENTER 4 Select the L...

Page 25: ...tion date Identification of stored measured values Therefore check the time at regular intervals Note After a fall of the supply voltage empty batteries or accumulator pack the date and time are reset...

Page 26: ...numerals with the number keys opens up 4 Enter the time using the number keys The digit to be changed is displayed underlined 5 Confirm the setting with START ENTER The time is set 6 Set the current D...

Page 27: ...te Delete i 4 of 1000 occupied i Filter No filter Display Illumination Auto off On Off Contrast 0 100 Brightness 0 100 Reset Interface Baud rate 1200 2400 4800 9600 19200 Output format ASCII CSV Conti...

Page 28: ...Settings functions The settings are given in the menu Configuration System Move to the Configuration menu with the MENU key Menu item Setting Description Language Deutsch English Fran ais Espa ol Sel...

Page 29: ...Data filter Erase all stored measurement datsets Delete Information on the number of occupied memory locations The settings are given in the menu Configuration System Measured value memory Move to the...

Page 30: ...e with PRT Quit the display with ESC RS232 download Downloads to the interface all measurement datasets that correspond to the filter settings The download is ordered according to the date and time Th...

Page 31: ...tput format Output format The settings are given in the menu Configuration System Interface Move to the Configuration menu with the MENU key Settings Menu item Setting Description Illumination Auto of...

Page 32: ...ate Date Format of the date display Date format The settings are given in the menu Configuration System Continue Date time Move to the Configuration menu with the MENU key Settings Menu item Setting D...

Page 33: ...ctional differences in their light transmittance Therefore if you want to achieve accurate and reproducible measurement results it is necessary to always index the sample cells and cells for calibrati...

Page 34: ...y if the cell is closed with the black light protection cap WTW cells Note The outside of the cell always has to be clean dry and free of fingerprints and scratches Clean the cells before starting to...

Page 35: ...the marking on the cell cap with the marking on the cell shaft Press and for a short time hold the START ENTER key until the measured value is displayed Unmarked cell see page 33 Press and hold the ST...

Page 36: ...calibration interval has expired With a temperature change Calibration procedures and calibration standards For the menu guided three point calibration you need the following three calibration standa...

Page 37: ...reparatory activities when you want to calibrate Note For turbidity values under 1 FNU NTU follow the instructions given in the appendix see APPENDIX 2 TURBIDITY VALUES UNDER 1 FNU NTU Carrying out ca...

Page 38: ...complete rotation by 360 After each step wait for a short time until the displayed measured value is stable Turn the cell back to the position with the lowest measured value 4 Release the START ENTER...

Page 39: ...layed as the calibration result a note appears on the display to recalibrate before measuring Should a valid calibration not be possible the meter also offers to continue measuring with the last valid...

Page 40: ...Measurement dataset A complete dataset consists of Date time ID number ID Measured value 4 6 1 Storing measurement datsets Proceed as follows to transmit to the data memory and simultaneously output...

Page 41: ...ll stored measurement datsets that correspond to the specified filter criteria The settings are given in the menu Configuration System Measured value memory Data filter Move to the Configuration menu...

Page 42: ...display of stored measurement datasets you have the following options Switch directly to the measured value display with M short pressure Leave the display and switch to the superordinate menu with ES...

Page 43: ...nerally a terminal program serves to establish a connection to a device on a data interface and to communicate with the device via a console on the display A terminal program usually offers the possib...

Page 44: ...t assignment 4 7 2 Configuring the RS232 interface For error free data transmission the RS232 interface should be set to the same transmission speed Baud rate on the Turb 430 IR T and PC printer You c...

Page 45: ...u Configuration System Interface Output format The ASCII output format delivers formatted datasets The CSV output format delivers datasets separated by Output format ASCII Turb 430 IR Ser no 12345678...

Page 46: ...measuring range limit only with measured value status OFL UFL 5 Unit of 4 Unit of the measured value 6 Measured value status of 4 VALID Measured value valid INVALID Measured value invalid UFL Measured...

Page 47: ...that is being shown on the display to the interface displayed measured values stored measurement datsets calibration record Data Operation description Current measured value Press PRT Simultaneously...

Page 48: ...ttable settings are reset Settings for Turbidity see section 4 8 2 System settings 4 8 2 Resetting turbidimeter settings With the Turbidity Reset function all turbidimeter settings are reset System se...

Page 49: ...version Series number of the meter 4 10 Software update With a software update you obtain the current instrument software see appendix The current software version can be found on the Internet under w...

Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...

Page 51: ...ION Make sure that the poles of the batteries are the right way round The signs on the batteries must correspond to the signs in the battery compartment 1 Open the battery compartment Unscrew the two...

Page 52: ...ection 8 1 1 Open the battery compartment Unscrew the two screws 1 on the underside of the meter Remove the lid of the battery compartment 2 2 If necessary take four old batteries out of the battery c...

Page 53: ...shaft If liquid is in the cell shaft e g due to a spilled cell clean the cell shaft as follows 5 2 2 Cleaning the cells Cells have to be clean dry and free of fingerprints Therefore clean them regula...

Page 54: ...st damage during transport Batteries accumulator pack Remove the batteries or accumulator pack from the meter see section 5 1 Dispose of the batteries or accumulator pack at a suitable facility accord...

Page 55: ...arge the accumulator pack see section 3 2 Cause Remedy Software error Operating condition undefined or EMC load unallowed Processor reset Press the START ENTER and PRT key simultaneously Cause Remedy...

Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...

Page 57: ...eries 4 x 1 5 V type AA Operating time with battery operation Turb 430 IR approx 3000 measurements Turb 430 T approx 2000 measurements Accumulator pack optional 5 x 1 2 V nickel metal hydride NiMH typ...

Page 58: ...gital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment Thi...

Page 59: ...red value or 0 01 FNU NTU Reproducibility 0 5 of the measured value Response time 4 seconds Calibration Automatic 3 point calibration Measuring principle Nephelometric measurement according to US EPA...

Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...

Page 61: ...et 251 302 Calibration standard kit for Turb 430 IR Kal Kit Turb 430 IR 600 560 Calibration standard kit for Turb 430 T Kal Kit Turb 430 T 600 561 Thermoprinter a connection cable is required to conne...

Page 62: ...540 S 902 843 Connection P3001 LabStation Cable AK 3000 250 745 in conjunction with an adapter socket socket GenderChanger Specialist shops or Cable 2 x 9 pin socket plug Specialist shops Description...

Page 63: ...n aids Abbreviations The list of abbreviations explains the indicators and the abbreviations that appear on the display and in the manual Specialist terms The glossary briefly explains the meaning of...

Page 64: ...T 64 ba75507e05 07 2016 Abbreviations Cal Calibration d Day h Hour j Year LoBat Batteries almost empty Low battery m Month s Second S Slope internat k SELV Safety Extra Low Voltage Slp Slope determine...

Page 65: ...ation Measured value The measured value is the special value of a measured parameter to be determined It is given as a combination of the numerical value and unit e g 3 m 0 5 s 5 2 A 373 15 K Measurin...

Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...

Page 67: ...7 31 Display illumination 7 20 F Filter 41 Firmware update 69 I Initial commissioning 17 Initialization 48 Inserting a cell 20 Interface 31 J Jack field 7 K Keys 6 M Measured value display 21 Measure...

Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...

Page 69: ...menu The program automatically selects the first free serial interface COM port The selected interface is displayed on the left side of the status line on the screen bottom Via the language menu you c...

Page 70: ...oceed as follows to measure turbidity values under 1 FU NTU Calibration procedure Calibrate after the calibration interval has expired with a temperature change Measuring 1 Press the CAL ZERO key The...

Page 71: ...Turb 430 IR T ba75507e05 07 2016 71...

Page 72: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...