Parameterization
98
With
set switching delay time.
Setting range: 0 ... 20 seconds.
With
accept and go on to
"MESSAGE OFF/MESSAGE ON".
Note:
A range of 0 ... 20 seconds can be selected as delay time (step size 1 s).
It works as slow operation and slow release.
With
select between ON and OFF.
With
accept and go on to next relay or to
submenu "PARA TIMER" (with selected timer
function for a relay - configuration: menu
CONF RELAYS).
Input of temperature limit values
With the selection of relay points as limit contacts for the temperature (configuration level:
submenu "CONF RELAYS") the submenus "PARA UL TP" or "PARA LL TP" for the respective
relay points are displayed in the submenu "PARA UL/LL" when selecting with the
keys.
Example: Relay point 4 has been configured as upper limit contact for the temperature.
Setting:
Submenu "PARA UL/LL".
With
enter submenu
"PARA UL/LL".