Queries / Settings
110
Extreme values in the course of the day
Between 0:00 and 24:00 the extreme values of the measured conductivity values are stored with
the corresponding time. The values of the 3 previous days remain stored and can be called up.
To eliminate short peaks, extreme values are formed as mean values over a selectable period.
The period for forming the mean values can be set to 1, 5, or 15 minutes. The values are
displayed in mS/cm or µS/cm corresponding to their form in the measuring mode.
Daily minimum and maximum values conductivity
Submenu "DAY MIN/MAX".
With
enter the "DAY MIN/MAX" menu.
Extreme values of the day displayed.
With
display extreme values.
Time max. value (on the left): e. g. 12:05
Time min. value (on the right): e. g. 22:55
Maximum (big indication): e. g. 6.06 mS/cm.
Minimum (small indication): e.g. 3.57 mS/cm.
With
select extreme values of the last
previous day.
With
display extreme values.