Technical data
118
AVERAGING
Averaging
Computing of daily MIN and daily MAX values out of mean
values that have been determined over a period of 1/5/15
minutes. The last valid daily minimum and maximum values with
the time when they occurred are stored. Storage in the
EEPROM.
The daily minimum and maximum values can be displayed.
SALINITY
Measuring rate
1 measurement/s
Measuring range
0.0 to 70.0
Resolution
0.1
CONDUCTIVITY CHECK
Aim
Check of the cell constant with 0.01 molar potassium chloride
solution
TEMPERATURE MEASUREMENT
Measuring rate:
1 measurement/s
Measuring ranges:
-5.0°C ... +60.0°C
Accuracy:
± 0.2 K ±1 digit
Resolution:
0.1 K
Additional fine adjustment by ± 0.5 K designed in the software in order to avoid display
discrepancies when changing the measuring cell.
TEMPERATURE COMPENSATION
The temperature probe is integrated in the measuring cell.
l
freely adjustable linear temperature coefficient
from 0.00 to 3.00 %/K, or
l
non-linear function "nLF" of natural water
according to DIN 38404
REFERENCE TEMPERATURE
20°C or 25°C switchable
in operation mode SAL the conversion refers to a reference
temperature of 15°C.