Configuration
84
609
Submenu conductivity measuring cell
TetraCon 300.
The cell constant (0. 609) is displayed.
With
selection of submenus.
With
go on to CONF LF/SAL
Submenu VAR.
The cell constant of theTetraCon 700
(0.917) is displayed as default value.
With
selection of submenus.
With
go on to CONF LF/SAL
(Setting of the cell constant in the range
0.450 ... 1.300 in the Check function.)
Selection of operation mode Conductivity/Salinity
(ST / R / RS models)
Selection of operation mode.
Chart:
Submenu
Setting possibilities
CONF LF/SAL
LF *)
SAL
*) Ex-factory condition
In the CONF LF/SAL menu the following operation modes can be selected:
l
LF
Conductivity measurement
l
SAL
Salinity measurement
(TREF 15°C fixed, measuring range 0.0 - 70.0 fixed)
Starting position: CONF LF/SAL
Starting menu: LF/SAL