Multi 3430
Operation
45
ba75865e02
12/2009
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
Note
You can display the slope in the units, mV/pH or % (see page 39).
Stability control
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).
Calibration record
The new calibration values are displayed when a calibration procedure
is completed.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see page 54).
Subsequently, you can transmit the displayed calibration data to the
interface, e. g. to a PC, with the
<PRT>
key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Determined val-
ues
Displayed calibration data
1-point
Asy
z
Zero point =
Asy
z
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
z
Zero point =
Asy
z
Slope =
Slp.
3-point to
5-point
Asy
Slp.
z
Zero point =
Asy
z
Slope =
Slp.
The calibration line is calculated
by linear regression.
Summary of Contents for Multi 3430
Page 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Page 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Page 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Page 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Page 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Page 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Page 101: ......