Operation
Multi 3430
50
ba75865e02
12/2009
Continuing with three-
point to five-point
calibration
(
Buffer
TEC)
10
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<MENU/ENTER>
.
The calibration display for the next buffer appears.
11
Finish the calibration procedure as a two-point calibration with
<M>
.
The calibration record is displayed.
30.09.2009 08:00
30.09.2009 08:00
12
Thoroughly rinse the IDS pH sensor with distilled water.
13
Immerse the IDS pH sensor in buffer solution 3.
14
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.