Operation
Multi 3430
68
ba75865e02
12/2009
Measurement
/
Measuring cell
/
Man. cell const.
0.450 ...
0.475
...
0.500 cm
-1
Display and setting option of the
cell constant of any measuring
cells (
man
).
Measurement
/
Temp. comp.
(TC)
/
Method
nLF
lin
Off
Procedure for temperature com-
pensation (see section 4.8.3).
This setting is only available for the
measured parameters,
ӕ
and
ρ
.
Measurement
/
Temp. comp.
(TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear tempera-
ture compensation.
This setting is only available when
the linear temperature compensa-
tion is set.
Measurement
/
Temp. comp.
(TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ӕ
and
ρ
.
Measurement
/
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
4.11.1).
Menu item
Possible
setting
Description