Operation
Multi 3430
66
ba75865e02
12/2009
4.8.3
Temperature compensation
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20
or
Tr25
.
You can select one of the following temperature compensation meth-
ods:
z
Nonlinear temperature compensation
(
nLF
) according to EN 27
888
z
Linear temperature compensation
(lin)
with adjustable coeffi-
cients of 0.000 ... 3.000 %
/
K
z
No temperature compensation (off)
Note
The reference temperature and temperature compensation are set in
the
Measurement
menu for the parameter, conductivity (see section
Application tips
Select the following temperature compensations given in the table
according to the respective test sample:
Test sample
Temperature compensation
Display
indica-
tor
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
lin
Set linear
temperature coefficient
0.001 ... 3.000 %/K
lin
Salinity (seawater)
Automatic
nLF
according to
IOT
Sal
,
nLF
Summary of Contents for Multi 3430
Page 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Page 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Page 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Page 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Page 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Page 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Page 101: ......