What to do if...
Multi 3430
92
ba75865e02
12/2009
6.3
Conductivity
Error message,
OFL
Error message,
Error
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable IDS conductiv-
ity sensor
Cause
Remedy
– IDS conductivity sensor contam-
inated
– Clean and if necessary,
exchange IDS conductivity
sensor
– Calibration solution not suitable
– Check calibration solutions