Multi 3430
Operation
69
ba75865e02
12/2009
Setting menu of
LR325/01
Menu item
Possible set-
ting
Description
Measurement
/
Measuring cell
/
Cell constant
0.090 ... ...
0.100
...
0.110 cm
-1
Display and setting option of
the cell constant of any mea-
suring cells (
man
).
Measurement
/
Temp. comp. (TC)
/
Method
nLF
lin
Off
Procedure for temperature
compensation (see section
4.8.3).
This setting is only available
for the measured parame-
ters,
ӕ
and
ρ
.
Measurement
/
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear tem-
perature compensation.
This setting is only available
when the linear temperature
compensation is set.
Measurement
/
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured parame-
ters,
ӕ
and
ρ
.
Measurement
/
TDS
factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.11.1).
Summary of Contents for Multi 3430
Page 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Page 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Page 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Page 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Page 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Page 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Page 101: ......