Operation
Multi 3430
84
ba75865e02
12/2009
ORP
The following settings for ORP measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required measured parameter in the measured value display and
press the
<MENU/ENTER>
key.
Conductivity
The following settings for conductivity measurements are reset to the
default settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required measured parameter in the measured value display and
press the
<MENU/ENTER>
key.
Setting
Default settings
Resolution mV
0.1
Man. temperature
25 °C
Stability control
On
Setting
Default settings
Cal. interval
150 d
Measured parameter
ӕ
Cell constant (c)
Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of
the linear temperature compen-
sation
2.000 %/K
TDS factor
1.00
Stability control
On
Summary of Contents for Multi 3430
Page 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Page 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Page 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Page 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Page 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Page 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Page 101: ......