Multi 3430
Operation
55
ba75865e02
12/2009
Sample printout
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
-
Displays the calibration
records.
Further options:
z
Scroll through the cali-
bration records with
<
W
><
X
>
.
z
Output the displayed
calibration record to the
interface with
<PRT>
.
z
Output all calibration
records to the interface
with
<PRT
_
>
.
z
Quit the display with
<ESC>
or
<MENU/
ENTER>
.
z
Switch directly to the
measured value display
with
<M>
.
31.07.2009 16:55
Multi 3430
Ser. no. 08502113
CALIBRATIONpH
SenTix 940
Ser. no. B211371282
31.07.2009 16:13:33
AutoCal TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage
1
184.0 mV 24.0 °C
Voltage
2
3.0 mV
24.0 °C
Voltage 3
-177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...