62
10 Supplement
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Temperature derating
1
2
0 °C
(32 °F)
-40 °C
(-104 °F)
55 °C
(131 °F)
80 °C
(176 °F)
130 °C
(266 °F)
100 °C
(212 °F)
-40 °C
(-104 °F)
Fig. 40: Temperature derating CPT-2x, version up to +130 °C (+266 °F)
1 Process temperature
2 Ambient temperature
1
2
0 °C
(32 °F)
-40 °C
(-104 °F)
50 °C
(122 °F)
80 °C
(176 °F)
150 °C
(302 °F)
100 °C
(212 °F)
-40 °C
(-104 °F)
Fig. 41: Temperature derating CPT-2x, version up to +150 °C (+302 °F)
1 Process temperature
2 Ambient temperature
SIP process temperature (SIP = Sterilization in place)
Applies to instruments configurations suitable for vapour, i.e. material measuring cell seal EPDM or
FFKM (Perlast G75S).
Vapour stratification up to 2 h
+150 °C (+302 °F)
Process pressure
Permissible process pressure
see specification "
process pressure
" on the type label
Mechanical stress
9)
Vibration resistance
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Shock resistance
50 g, 2.3 ms according to EN 60068-2-27 (mechanical
shock)
10)
9)
Depending on the instrument version.
10)
2 g with housing version stainless steel double chamber