8. Remote Operation
62
WIKA Operating Instructions High Pressure Controller, Model CPC8000-H
GB
1
4
2
3
5
9
9
3
.0
0
/0
6
/2
0
1
7
/GB
Item Command or
Query
Data
Response/Function
65.
LISTRANGE?
PRI,1, min,max
Returns the range of the
installed sensor for the
active channel
66.
LOCALE
locale code
Sets Language and
Country code.
Locale
Code
Languag
e
Country
en_us
english
usa
en_ca
english
canada
en_gb
english
Great britten
fr_fr
french
france
fr_ca
french
canada
es_es
spanish
spain
es_mx
spanish
mexico
de_de
german
germany
it_it
italian
italy
zh_ch
chinese
china
ko_ko
korean
korea
ru_ru
russian
russia
de_ch
german
switzerland
fr_ch
french
switzerland
en_ie
english
ireland
ja_jp
japanese
japan
pl_pl
polish
poland
pt_pt
portugue
se
portugal,
brazil
67.
LOCALE?
<sp>{string}<cr><lf>
Returns current language and
Country locale.
68.
LOCALGRAVITY
value in ft/s^2
Sets the local gravity in feet/sec^2.
69.
LOCALGRAVITY?
<sp>n.nn<cr><lf>
Returns the local gravity in
feet/sec^2
70.
LOWERLIMIT
Value inside primary
xducer in current units.
Sets the lower control limit for the
instrument
71.
LOWERLIMIT?
<sp>n.nn<cr><lf> Returns the lower control limit for
the instrument in current units.
72.
MEASURE
none
Instrument placed in Measure
Mode
73.
MEASURE?
<sp>(YES or No)<cr><lf>
Returns Yes if instrument is in
measure. No if otherwise
74.
MODE
Standby, Measure,
Control,
Vent
Sets the operation mode
75.
MODE?
<sp>XXXXXX<cr><lf>
Returns the operation mode
76.
NETMASK
nnn.nnn.nnn.nnn
Sets the Ethernet network mask.
77.
NETMASK?
<sp>nnn.nnn.nnn.nnn
<cr><lf>
Gets the Ethernet network mask.
78.
OUTFORM
1 to 7 - Table 5
Sets the output format.