8. Remote Operation
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
57
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
Separator
Space (SP).
Data
ASCII representations of numbers, {value}, or alpha characters, {string}, data as defined
above. When sending code a literal variable replaces the brackets and the enclosed
character(s) shown in the following examples.
Termination
Linefeed (LF) or carriage return (CR) is used to signal the end of a command statement. For
IEEE-
488.2 operation “EOI” is an acceptable alternative. Always send commands in one of the
following formats:
1. [Command] [Termination];
2. [Command] [Separator] [Data] [Termination];
3. Queries are special instructions in the form: [Command?] [Termination] where the
question mark, “?”, immediately precedes the terminator.
When a valid query is received, the CPC8000-H will return {data} terminated by CR and LF.
Floating point data is returned in the current engineering units in exponential format.
Output Formats
Pressure readings are returned in exponential notation in a format according to the
OUTFORM command as follows. Outform applies to both pressure channels.
Table 5 Output Formats
Item
Format
1.
<sp> pressure value <cr><lf>
2.
<sp> pressure, units number,STANDBY <cr><lf>
3.
<sp> pressure, pressure rate <cr><lf>
4.
<sp> pressure, minimum peak, maximum peak <cr><lf>