7. Operation
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
35
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
Figure 7-10 Secondary and Tertiary Displays
Figure 7-11 Screen of Display Choices
7.2.12 Limits
Limits are shown in three places in the main screen. The “Range Limits”, “Control Limits” and “Entry
Limits” shown in “
below can show two different readings depending on settings.
The Range Limits shows the limits of the range selected in Range Hold mode or the limits of the active
transducer in Autorange mo
de. The “Control Limits” are the limits entered into the Setup [
] /
[Control] [Maximum Limits]/[Minimum Limits] screen (see Section 7.3.3.1). The
“Control Limits” limit
the value that can be entered as a setpoint. The “Entry limits” are equal to the
lessor of the Control
Limits or the Range Limits (when in Range Hold mode).
Figure 7-12 Limits
7.2.13 Bar Graph
The bar graph is always active and reports in real time the pressure at the Measure/Control port. This
provides a visual comparison between the actual pressure and the user selected control limits.
Secondary Display
Tertiary Display
Range
Limits
Control
Limits
Entry
Limits