8. Remote Operation
70
WIKA Operating Instructions High Pressure Controller, Model CPC8000-H
GB
1
4
2
3
5
9
9
3
.0
0
/0
6
/2
0
1
7
/GB
Item
Command or Query
Response/Function
SENSE
16.
[:PRESsure] [z]
Selects the sensor to read or modify or active sensor if [z]
is omitted.
17.
:NAME?
Returns the accuracy and range of the sensor.
18.
:MODE?
Returns pressure type “ABSOLUTE” or “GAUGE”
19.
:MODE ABS/GAUGE
Sets the pressure type to absolute or gauge.
20.
:ABS?
Returns native sensor type “1” for absolute, ”0” for gauge.
21.
:RESolution?
Returns the FS-related relative resolution of the sensor.
:RANGe
22.
[:UPPer]?
Returns the maximum range of the primary sensor.
23.
:LOWer?
Returns the minimum range of the primary sensor.
:Unit
24.
[:NAME]?
Returns a string representing the native units of the sensor.
25.
:VALue?
Returns the conversion factor of the native units of the
sensor to the current unit.
:REFerence
26.
[:HEIGht] <n>
Sets the head height correction [mm].
27.
:HEIGht?
Returns the head pressure height.
28.
:MODE?
Returns head correction mode -
“OFF”, “GAS”, LIQUIT”.
29.
:MODE
OFF/GAS/LIQUIT
Sets the head correction mode.
30.
:MEDIUM <n>
Sets the density of the medium at 20°C & 1bar(abs)
[kg/m3] (or equivalent value in current units).
31.
:MEDIUM?
Returns medium.
32.
:ACTive <n>
Sets the active sensor
33.
:ACTive?
Returns the active sensor
SYSTem
34.
:DATE <i,i,i>
Sets the system date (YYYY,MM,DD).
35.
:DATE?
Returns system date YYYY,MM,DD
36.
:TIME <i,i,i>
Sets the system time (hh,mm,ss).
37.
:TIME?
Returns system time HH,MM,SS.
38.
:ERRor[:NEXT]?
Return: error#, “description”.
39.
:KLOCk ON/OFF/1/0
Lock or unlock keyboard.
40.
:PRESet
Load default values
41.
:SAVe
Saves settings to non-volatile memory.
42.
:VERSion?
Return of the SCPI-standard.
TEST
43.
:ELECtronic?
Returns “OK”.
44.
:RELay<n>?
Returns whether relay 1 to 3 is closed or not.
45.
:RELay<n> ON/OFF
Turns relay 1 to 3 on or off.