500 Clear All Remote DLC scale allocations : calibration command
This command will reset and invalidate any scales allocated to scale channels 1, 2 and 3.
All information will be erased. This command is typically used for a new installation to
clear any existing scale allocations. This command is also recommended as the first
command to be executed during a new, first time configuration.
To perform a remote reset function, press 1 followed by the [TARE] key. The indicator
will clear all scale allocations locally and remotely inside the DLC slaves.
501 Clear Scale 1 Allocation: calibration command
502 Clear Scale 2 Allocation: calibration command
503 Clear Scale 3 Allocation: calibration command
This command will reset and invalidate the current calibrated scale. The M2000D has 3
scales, to enter calibration mode for any one, use commands 19, 29 or 39. All information
will be erased. This command is typically used for a new installation to clear any existing
scale allocations to scale 1.
To perform a remote clear function on scale 1, press 1 followed by the [TARE] key. The
indicator will clear the scale allocated locally and remotely inside the DLC slaves.
A
C
GENERAL DLC
COMMANDS
NOTE: All scale calibration data for all 3 scales will be cleared and erased. All data will be lost. This
command is of use typically in new installations where the installation and calibration process needs to
be performed.