The next example of hopper scale allocation assumes the following scenario:
q
We have two 8 channel DLC slaves for a total of 16 cells
q
We have both Primary and Secondary DLC units connected to the indicator
q
We require 3 hopper scales of 4 corners each
The M2000D supports 3 scale channels. We must allocate a single hopper scale per scale
channel. First we enter calibration mode for scale 1 by executing command 19, and after
entering a password we allocate the first hopper scale using calibration command 520.
After completing the first scale allocation, we exit calibration mode. The next step is to
allocate hopper scale 2 by entering calibration mode for scale 2, the executing command
29. After entering a password we allocate hopper scale 2 using calibration command 520.
Once we complete the hopper scale installation, we exit calibration mode. The final step is
to enter calibration mode executing command 39 to allocate the third hopper scale. Once
in calibration mode we use calibration command 520 to allocate the third hopper scale.
Once all 3 the hopper scales are allocated based on the above criteria using calibration
command 520, we may observe the following allocation setup information:
SCALE 1
SCALE 2
SCALE 3