Electrical interference immunity
57
Fast transient
electrical
bursts in accor-
dance with IEC
61000-4-4
± 2 kV for
power supply
lines
±1 kV for
input/output
cables
Not
applicable
This medical electrical
device runs exclusive-
ly on battery power.
Electricity
surges
IEC 61000-4-5
± 0.5 kV,
± 1 kV
cable(s) to
cables
± 0.5 kV,
± 1 kV,
± 2 kV
cable(s) to
ground
Not
applicable
Power supply
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
Confor-
mant
Current frequency
magnetic fields
should correspond to
the characteristics of
a typical location in a
typical commercial or
hospital environment.
Summary of Contents for 96373
Page 3: ...A 3 1 2 3 4 5 ...
Page 4: ...B C 4 14 13 12 11 6 7 8 9 10 15 16 ...
Page 5: ...D 5 E F 2 28 28 27 26 25 24 17 18 19 20 21 23 22 ...
Page 46: ...Adhesive electrodes 46 Example for application of adhesive electrodes 1 57 1 57 in 40 40 mm ...
Page 109: ...Electrodos adhesivos 109 Ejemplos de uso de electrodos adhesivos de 1 57 1 57 in 40 40 mm ...