Electrical interference immunity
55
Guidelines and manufacturer’s declaration – electromag-
netic interference immunity
This device is intended for use in electromagnetic envi-
ronments as described below. The customer or operator is
responsible for observing that the device is used in such an
environment.
Interference
immunity
tests
Test level
in accor-
dance with
IEC 60601
Degree
of con-
formity
Electromagnetic
environment
Electrostatic
discharge
(ESD) in accor-
dance with IEC
61000-4-2
± 8 kV
contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
Confor-
mant
Floor surfaces should
be wood, concrete
or ceramic tile. If the
floor has a synthetic
covering, the relative
humidity must be at
least 30 %.