4 Maintenance procedures
This section provides the maintenance guidelines and procedures necessary to maintain the
mass spectrometer's performance.
Keep to a maintenance schedule, and perform maintenance as required and described in this
section.
4.1 Maintenance schedule for Xevo TQ-S cronos
The following table lists periodic maintenance schedules that ensure optimum instrument
performance.
Table 4–1: Maintenance schedule
Procedure
Frequency
For information...
Clean the instrument case.
As required.
See
Empty the nitrogen exhaust
trap bottle.
Check daily, empty as
required.
See
.
Inspect and adjust the
roughing pump oil level.
Weekly.
See
Clean the source components. When source components are
visibly fouled, the background
or high-peak contaminants are
unacceptably high, or
sensitivity decreases to
unacceptable levels.
See
Replace the ESI probe tip.
When sensitivity decreases to
unacceptable levels, or if
blocked or damaged.
See
Replace the probe assembly.
When sensitivity decreases to
unacceptable levels or signal is
unstable due to inconsistent
sample flow.
See:
Clean the APCI probe tip.
(Options using the APCI probe
adapter only.)
When sensitivity decreases to
unacceptable levels or when
significant chemical
interference is present.
See
.