a. Use the wash bottle containing 1:1 methanol/water to rinse the component over the
large beaker.
b. Blow dry the component with inert, oil-free gas.
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
7. Inspect each component for persisting contamination.
Requirement:
If contamination is present, clean the component again. If contamination
is still present, dispose of the component, according to local environmental regulations,
and obtain a new one before reassembling the sampling cone assembly.
4.12.4 Assembling the sampling cone assembly
Required materials
Chemical-resistant, powder-free gloves
To assemble the sampling cone assembly:
Notice:
To avoid recontaminating the components, wear clean, chemical-resistant,
powder-free gloves.
Notice:
To avoid damaging the sampling cone, which is fragile, do not place it on its
tip. Always place it on its flanged base.
1. Carefully fit the sample cone into the cone gas nozzle.
Figure 4–14: Sampling cone assembly
O-ring
Sample cone
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Page 94