D Plumbing the fluidics system
This section provides reference information for replacing the tubing in the fluidics system.
Warning:
To avoid electric shock, do not use stainless steel tubing or stainless steel
finger-tight screws to connect the selector valve to the source probes; use the PEEK
tubing and natural (beige) colored PEEK finger-tight screws.
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
D.1 Preventing contamination
For information on preventing contamination, refer to
Controlling Contamination in LC/MS
Systems
(715001307). You can find this document on
Support
>
Support Documents and Downloads
.
D.2 The selector valve
The selector valve is located behind the visor on the right-hand side of the instrument. The letters
etched on the front of the valve represent the components to which the associated port is
connected. Etched letters with a light background indicate inputs, while letters with a dark, etched
background indicate outputs. The following table outlines the component to which each port
connects, and whether the port is a fluid input or output.
Table D–1: Selector valve connections
Port
Component
Input/Output
P
Infusion pump
Both
A
Sample reservoir A
Input
B
Sample reservoir B
Input
R
Rinse/wash bottle
Input
W
Waste bottle
Output
LC
LC column
Input
S
Xevo TQ-S cronos ion source
Output