the correct probe assembly. For example, the UniSpray probe assembly will not fit the tool-
free probe adapter.
4. Tilt the probe adapter cap so that the ball bearing is located in the recess at the bottom of
the aperture, and then insert the probe assembly tubing through the aperture.
Figure 3–14: Probe adapter cap
Probe cap aperture from the underside
Probe cap aperture from the top
Tip:
The probe assembly tubing can pass through the aperture only when the ball bearing
is located in the recess at the bottom and does not block the aperture. To ensure that the
probe assembly does not contact the high voltage inside the probe adapter, the ball
bearing prevents you from installing the probe assembly when the probe adapter is fitted to
the source enclosure.