Figure 4–36: PEEK fitting
PEEK fitting
3. Remove the probe assembly.
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
4. Dispose of the probe assembly in accordance with local environmental regulations.
5. To install a new probe assembly, see
Installing the probe assembly (Page 54)
4.17 Replacing the ESI probe tip and gasket
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing through which the
stainless steel capillary passes or if the probe tip is damaged.
4.17.1 Removing the ESI probe tip and gasket
Required materials
• Chemical-resistant, powder-free gloves
• 7-mm open-end wrench
• 10-mm open-end wrench
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Page 122