June 11, 2015, 715004754 Rev. A
Page 96
Figure 6–2:
Sides of spring-loaded fuse holder
3. With minimum pressure, withdraw the spring-loaded fuse holder.
Figure 6–3:
Removing fuse holder
4. Remove and discard the old fuse.
5. Ensure that the new fuse is properly rated for your requirement.
6. Insert the new fuse into the fuse holder.
7. Insert the fuse holder into its receptacle in the detector’s rear panel, and gently
push on the fuse until it locks into position.
Warning:
To avoid fire, replace the fuse only with another of the same
type and rating.
Table 6–1:
Fuse requirements
Nominal voltage
Frequency
Required fuse
100 Vac to 240 Vac
50 to 60 Hz
F 3.15 A, 250 V
Sides of fuse holder
Fuse holder
Power-entry module
Fuse (2)