Note:
When the detector is operating under 410 mode, the method icon does not
appear. Alternatively, if the detector is controlled by the system controller of a data
system, the remote icon is displayed.
The detector method includes all of the parameter settings necessary to reproduce a
chromatographic separation.
5.2.4.1
Timed events
You can program as many as 16 timed events, to the nearest 0.01 minute. As you
enter timed events, each new event is appended to the end of the timed event list. If
you enter a time that is not in sequence with the events entered previously, the timed
event list is sorted automatically when you press Next. The 2414 RI detector allows
programming of the timed events shown in the following table.
To program a new timed event:
1. On the keypad, press the METHOD (Shift TEMP
°
C) key.
Figure 5–7:
Method choice list screen
Table 5–4:
Timed event parameter
Number
Event
Units
Range
Default
1
Autozero
Does not apply
Does not apply
Disabled
2
Polarity
Does not apply
1. –
+
2. +
3
RIU-FS
µ
RIU
1 to 500
Disabled
4
Chart mark
Does not apply
Does not apply
Disabled
5
Start Purge
Does not apply
Does not apply
Disabled
6
End Purge
Does not apply
Does not apply
Disabled
7
Switch
µ
RIU
1. On
Disabled
2. Off
3. Pulse
4. Rectangular
Wave
5. Toggle
8
Threshold
µ
RIU
0.001 to 500.000 Disabled
9
Start Recycle
Does not apply
Does not apply
Disabled
10
End Recycle
Does not apply
Does not apply
Disabled
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...