June 11, 2015, 715004754 Rev. A
Page 50
Figure 3–6:
Chart mark connections between the Alliance e2695
separations module and the 2414 RI detector
3.4.1.3
Generating chart mark and autozero signals
To generate both a chart mark and an autozero signal from the Alliance e2695
separations module to the 2414 RI detector, make the connections shown in the table
and figure below.
Table 3–5:
2414 RI detector connections to an Alliance e2695 separations
module
Alliance e2695 separations module
(connector B)
Detector (connector A)
Pin 1 Inject Start
Pin 9 Chart Mark +
Pin 2 Inject Start
Pin 10 Chart Mark –
Table 3–6:
Detector connections to an Alliance e2695 separations module
Alliance e2695
separations module
(connector B)
Detector (connector B)
Detector (connector A)
Pin 1 Inject Start
Pin 1 Autozero
+
Pin 9 Chart Mark
+
Pin 2 Inject Start
Pin 2 Autozero –
Pin 10 Chart Mark –
Red
Black
Inject Start
Inject Start
Ground
Stop Flow+
Stop Flow–
Hold Inject 1+
Hold Inject 1–
Hold Inject 2+
Hold Inject 2–
Ground
Chart Out+
Chart Out–
1
2
3
4
5
6
7
8
9
10
11
12
1 + Inject Start
2 - Inject Start
3 Ground
4 + Purge
5 - Purge
6 + Polarity
7 - Polarity
8 Ground
9 + Chart Mark
10 - Chart Mark
11 Ground
12 + Switch
13 - Switch
2414 RI detector connector A
Alliance e2695
separations module
connector B
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...