June 11, 2015, 715004754 Rev. A
Page 55
3.4.4 Polarity connections
The Polarity and Polarity Enable contact closures on the rear panel of the 2414 RI
detector determine the peak polarity of the output signal according to the following
conditions (negative polarity results in negative, or inverted, peaks):
•
Polarity serves as a positive/negative input.
•
Polarity Enable serves as an external input (Polarity) enable.
•
When Polarity Enable is open (not connected), the polarity is determined from
the polarity selected on the front panel or, if chromatography data software is
used, from the polarity method setting.
•
When Polarity Enable is closed (connected to an instrument), polarity
determines peak polarity. Polarity open (disconnected) generates negative
polarity. Polarity closed (connected) generates positive polarity.
The following table summarizes the polarity options:
3.5 Connecting to an external column heater
The 2414 RI detector can control one optional, external, column heater through the
EXT 1 port (external column heater port) on its rear panel. The port is a standard
9-pin DIN connector.
Figure 3–11: Location of external column heater port on 2414 RI
detector rear panel
Table 3–10: Polarity connection options
Polarity enable
Polarity
Detector out polarity
Open
Open
No Effect
Open
Closed
No Effect
Closed
Open
Negative (Inverted)
Closed
Closed
Unchanged
External
column
heater port
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...