June 11, 2015, 715004754 Rev. A
Page 100
7.2.2 Using the Optimize LED diagnostic function
To use the Optimize LED diagnostic function, press DIAG 2.
This diagnostic function displays the type of LED used (and automatically sets the
energy level of the internal LED. An error message appears if the energy level is too
high or too low. This message commonly appears when the detector was purged
recently or is about to begin a new analysis, or when a mobile phase composition
change affects the flow cell. The LED level can change if the temperature of the
internal oven changes.
Figure 7–4:
Optimize LED screen
7.2.3 Using the input and output diagnostic functions
Use the input and output diagnostic functions to perform these tasks:
•
Fix (that is, specify) voltage
•
Monitor event-out switch and eight contact closure event inputs
To access either of the input and output diagnostic functions, press DIAG 3, Input &
output.
Figure 7–5:
Input & output diagnostics screen
Other Diagnostics
Miscellaneous tests and functions revealing the view
autozero offset, oven and external column heater
configuration and status, generate test peaks, check purge
and recycle valve status and operation.
1. Autozero offset
2. Oven/column heater
3. Simulate Peak
4. Valve
5. Previous choices
Service Diagnostics
Diagnostics used by Waters service personnel.
Table 7–1:
Detector diagnostics (continued)
Diagnostic
Description
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...