Maintenance
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
85
400-cycle maintenance
REPLACING THE BACTERIOLOGICAL FILTER
A video animation is available on https://video.wh.com/en_global/dental-faq.
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 79.
Proceed as follows:
1
Open the service door.
2
Unscrew the bacteriological filter by hand (counter-clockwise).
3
Screw on the new bacteriological filter (clockwise) and tighten it snug.