Diagnostics
102
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
Code
Description
Actions
3xx
Check the door gasket. Clean or replace it if necessary.
Repeat the cycle.
If the problem persists, call service.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.
4xx
Clean water error (bad quality or low water level). Drain and refill the clean water tank.
Repeat the cycle.
If the problem persists, call service.
5xx
Check if there are hurdles on the door locking area (chamber rack, loads, objects, …).
Repeat the cycle.
If the problem persists, call service.
Check the door gasket (wrong placed).
Check if the door can move freely without touching the trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.
MESSAGES AND ALERTS
Notice: For any error not listed in this table, call technical service.
Message/Alert
Description
Action
Fill clean water tank.
There is not enough water in the tank to perform a
cycle.
Fill the water tank as requested.
Drain used water tank.
The used water tank is full.
Drain the water tank as requested.
Please close the door.
The door must be locked, but you didn't close it.
Close the door so it can be locked.