Sterilizer tests
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
53
4
Place the tube with the capsule in the center of a tray in the lowest rack position
and close the chamber door.
5
On the homepage tap Helix-B&D test.
6
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
7
Tap
and enter your credentials if required: the chamber door locks.
8
Wait until the end of the test and tap OPEN: the chamber door unlocks.
9
Enter your credentials if required.