Sterilization cycles
STOP A STERILIZATION CYCLE
WARNING! You can stop the cycle at any time. Instruments
must not be considered sterile if this occurs before the DRY
phase.
A cycle can be manually aborted at any time. To stop a cycle:
1
Tap STOP: a confirmation request
appears.
2
Tap
to abort the stop command.
The cycle continues programmed.
3
Tap
to abort the cycle: the
sterilizer starts a reset phase.
Notice: Do not switch off the sterilizer
during the reset phase: it takes some
time to reset the system and reach
safe conditions in the sterilizer
chamber.