Sterilization cycles
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
57
CONTENTS
This section deals with the following subjects:
Load maintenance and preparation
Sterilization cycle description
Sterilization cycle management
Load maintenance and preparation
WARNINGS
WARNING! Any residual of chemicals (like cleaning and
disinfection products), could affect the purity of the steam
and consequently the whole sterilization process.
If necessary, the load shall be cleaned and lubricated in
accordance with the instrument manufacturer’s
instructions.
Notice: Any residual of chemicals could seriously damage the
sterilizer. The manufacturer’s warranty is void in case of damage
caused by chemicals.
DENTAL HANDPIECES EXTERNAL DISINFECTION
This procedure reduces the risk of infection during cleaning and
maintenance of the dental handpieces.
n
Wear protective gloves during disinfection.
n
Avoid using abrasive disinfectants (pH-value 2.5 – 9; no
chlorine based disinfectants).
n
Use disinfectant wipes rather than spray disinfection.
n
Do not immerse handpieces in disinfectants.
n
Residual disinfectants on handpieces can cause extensive
damage to your instrumentation during sterilization (oxidation,
alteration of technical characteristics of seals, rubbers, fiber
optics, etc.)