Sterilizer tests
54
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
10
Open the chamber door, extract the tray using the tray holder and take the tube.
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle.
Wear appropriate PPE (e.g. gloves).
11
Unscrew the tube capsule and remove the indicator strip.
12
Check the change in colour. See "Interpret the test result" below
INTERPRET THE TEST RESULT
Indicator
What happened
Test
passed
What to do next
The indicator strip has turned dark. The air was completely removed from
the capsule.
Yes
-
Part of the chemical indicator strip has not turned dark. The air removal
from the capsule was not complete.
No
Repeat the test. If it fails repeatedly,
call technical service.
WHAT TO DO NEXT
Compile the Helix test documentation form to trace the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. See
"Helix test documentation form" on page 136.
Vacuum test
CAUTION! Follow local/national guidelines on the frequency of testing.
Notice: If a drainage period of the S Fast 134 cycle is still operating, wait the drainage to be finished and the sterilizer to be both cold and dry.
Otherwise, a false negative outcome could occur.