10. Operation
Change program
Press the ORANGE button to select programs 1 – 4 in ascending order. In program 5 switch the torque steps from
20 – 60 Ncm. The motor direction is automatically set to forward operation every time the program is changed.
When changing from program 4 to program 1 and in program 5 from 60 Ncm to 20 Ncm, a longer acknowledgment signal
sounds (risk of injury).
Pump ON / OFF
Only when the motor is at complete standstill can the pump be switched on or off by pressing the GREEN button of the foot control.
If the pump function is activated, the pump symbol is shown on the display.
Reverse operation
Press the YELLOW button to change from forward operation to reverse operation. On selecting reverse operation, an audible signal
can be heard and the selected program button flashes. Before the motor starts in reverse operation, 3 audible warning signals are
given.