36
12. Thread cutter function (chip breaker mode)
Press P5 program button.
Use Plus/Minus to increase or decrease the torque.
When the thread cutter function (P5) is activated, the speed in both forward and reverse operation modes is 20 rpm and
can no longer be changed.
When the motor button (grey) on the foot control is pressed, the thread cutter rotates inwards
until the set torque is reached. The control unit automatically switches to reverse operation when the set torque is reached.
Disengaging and then re-engaging the motor button will switch the control unit back to forward operation.
If the thread cutter function is in reverse operation mode, the control unit can also start with the maximum torque.
Summary of Contents for Implantmed
Page 1: ...Instructions for Use SI 915 SI 923...
Page 56: ......
Page 58: ......
Page 60: ......
Page 63: ......