Control unit operation
Program P4:
range 5 – 70 Ncm, intermediate stage 32 Ncm.
Program P5:
range 20 – 60 Ncm.
Pressing and holding PLUS / MINUS depressed activates the repeat function and the values are continuously
increased / decreased.
A longer confirmation signal is heard on changing from 5 to 70 Ncm (20 to 60) and 70 to 5 Ncm (60 to 20).
The motor switches off automatically when the set torque is reached in forward and reverse operation modes.
Press the program button (P4 or P5)
P4:
increase torque in 5-Ncm steps
P5:
increase torque in 10-Ncm steps
P4:
decrease torque in 5-Ncm steps
P5:
decrease torque in 10-Ncm steps
The accuracy of the set torque in the 20 – 50 Ncm range for the W&H WI-75 E/KM contra-angle handpiece is ± 10%.
Greater deviations may be encountered with other instruments.
Changing torque (P4 – P5)