| Handling and control
50
|
CO2 Incubator Basic
VWR
Chapter 7
Resetting the thermal protection
1. The Thermal Protection Active LED illuminates:
2. Turn the device off:
Press the power switch.
All indicators come off.
3. Turn the device back on:
Press the power switch.
NOTICE!
Thermal protection:
When the cause of the failure (e.g. excessive temperature in the operating room) has been
repaired, the device is set to normal incubation operation after it has been turned on again. If the
cause of the failure cannot be repaired with simple measures (e.g. by ventilating the room or by
reducing the temperature in the operating room), the thermal protection will respond again
immediately; in this case, contact Technical Service!